Infrared spectra matching and fingerprint identification
In the simplest case, two spectral printouts – one of the reference and the other of the analyte can be overlaid on a light box and the spectral features related by eye. Overlaying spectra on the computer screen achieves the same objective. When the spectrum of a substance being examined is compared with a reference spectrum, the positions and relative intensities of the absorption bands of the spectrum of the substance being examined should conform to those of the refer ence spectrum. When the two spectra are compared, care should be taken to allow for the possibility of differences in resolving power between the instrument on which the reference spectrum was prepared and the instrument used to examine the substance. It is good practice to run a spectrum of a polystyrene film on the same instrument to compare it with that recorded on the reference spectrum. The greatest variations through differences in resolving power are likely to occur in the region between 4000 and 2000 cm-1.
When a chemical reference substance is available, the substance being examined and the chemical reference substance should be prepared by the same procedure before recording the spectra (see later under Polymorphism). The transmission minima in the spectrum obtained with the substance being examined should correspond in position and relative size to those in the spectrum obtained with the reference substance.
In recent years, IR spectral compilations have been created and stored electronically in data bases and/or libraries. The spectrum of the analyte is presented to the database and the computer attempts to match the spectrum with one already held in the database. A report is made of the best matches. The computer program lists the most likely hits in order of a closeness of fit. Many spectra compilations (databases) are private collections, held typically by individual pharmaceutical companies; some can be purchased and a few are in the public domain.
The number of compounds for which IR spectra have been measured is now massive. Potentially, the greater the number of spectra in a database, the greater is the probability of making a good match for the unknown sample. However, the probability of making a mismatch is also greater, as more spectra with fine differences are available for comparison. The computer is simply matching ‘pictures’ by the number of peaks, their positions and their relative intensities. The best the computer fitting can do is to indicate a mathematical similarity. It is important to qualify a computer search:
• a visual overlay of the test compound spec trum and the hit spectrum ensures that the search has not chosen a match that is mathematically acceptable but not chemically acceptable
• knowledge of the class of a compound can help restrict the search to a more refined refer ence set (database) and can also help in deciding whether the substance given as a best fit is actually likely to be present in a sample
• other properties of the sample and the reference compound should match, such as chromatographic retention times, chemical and colour reactions, where these are available, and functional group assignments
• the computer can only select spectra that are in its library; if the spectrum of the compound under investigation is absent, then it will select those that give the next-best fit
• different forms of the same compound give different IR spectra (different polymorphs, racemate and/or enantiomer, ionisation status, cations and anions).
Some examples of the identification of drugs are given below.