Narcotic analgesics, opiates and opioids
المؤلف:
Sue Jickells , Adam Negrusz (Editors)
المصدر:
Clarkes Analytical Forensic Toxicology
الجزء والصفحة:
510
2026-11-28
8
Narcotic analgesics, opiates and opioids
Many laboratories perform specific assays for opiates for federal or legal purposes; these are generally limited to codeine, morphine and more recently 6-monoacetyl morphine (MAM; Paul et al. 1999). However, for clinical purposes a wider range of analytes is desirable and can include codeine, dihydrocodeine, hydrocodone, hydromorphone, oxycodone and oxymorphone. All assays involve a hydrolysis step (acidic or enzymatic; see p. 498 for an evaluation of these) to cleave the glucuronide conjugates, followed by a basic extraction (often using solid phase or acidic back extraction for cleanliness). Derivatisation is possible with a number of reagents (PFP, TMS, TFA or AC derivatives are the most common; Maurer and Pfleger 1984; Chen et al. 1990; Grinstead 1991). The derivatising reagent is selected on the basis of personal preference for a desired separation or the formation of unique ions on MS fragmentation. Analysis of hydromorphone, oxycodone and oxymorphone is complicated by the possibility that several structurally different derivatives will form in non-reproducible proportions from the tautomerisation of the enol and keto forms. However, these compounds can be stabilised in their keto forms by incubating with hydroxyl amine or methoxyamine–pyridine, and then yield only a single derivatised oxime product (Broussard et al. 1997; Meatherall 1999).