Mass spectrometry (MS) detector
Ion-trap mass spectrometer
In ion-trap mass spectrometers the production of ions in EI or chemical ionisation (CI) mode is achieved in pulses rather than continuously. The fundamental difference is that all the solute ions generated over the entire pulse period are trapped in the detector and are then sequentially ejected in increasing mass number from the trap into the electron multiplier. The addition of helium into the trap (133 mPa) contracts the ion trajectory to the centre of the trap, where it is further focused by the ring electrode to form dense ion packets that are expelled more efficiently than diffuse clouds, and thus greatly improves resolution. The spectral patterns can be quite different from those produced by mass filter spectrometers, and are often characteristic of the conditions under which the instrument is run, which makes comparison difficult between instruments. However, because the ion collection period is longer, the sensitivity of the ion trap in full scan mode is similar to that obtained in SIM on the average mass spectrometer. Furthermore, an improved mass range (sometimes up to several thousand a.m.u.) gives this type of detector many applications, particularly for quantitative trace analysis, and for higher mass components.